| Measurement of flatness, way straightness and small angles LS 128 A,F |
At the measurements of angles, way straightness and flatness the angle sensitive method is used.
| Measurement of small angles (LS 128A) |
Two optical units are used for this
measurement. The laser beam is split in the differential
beamsplitter LS 128.1 into two parallel beams which then enter
the reflector LS 128.2 with two corner cube reflectors. Reflected
beams return back into the differential beamsplitter. Reflector
LS 128.2 functions as the angular sensor. At the measurement the
differential beamsplitter LS 128.1 is fixed to the stationary
machine part and reflector LS 128.2 is fixed to the moving part.
Incidental inclination of the moving reflector evokes a path
difference between both split laser beams. This path can be
evaluated as inclination of the angular sensor.
![]() |
| Measurement of way straightness and flatness (LS 128F) |
For these measurements the angular sensor LS 128.2 is fixed on one of the three bases with footspacing 50, 100 or 150 mm according to the total assumed distance travelled. By moving the base with the angular sensor in steps along the particular base footspacing it is possible to obtain the required height profile in one direction (straightness). With the help of one or two auxiliary mirrors which bend the laser beam into the desired directions it is possible to obtain a complete three dimensional height profile of the measured table in different directions.
![]() |
| Set LS 128A | Set LS 128F | |||||
| Differential beamsplitter | LS 128.1 | Turning mirror | LS 128.3 | |||
| Angular sensor | LS 128.2 | Base 150 mm | LS 128.4 | |||
| Software | LS 128.7 | Base 100 mm | LS 128.5 | |||
| Base 50 mm | LS 128.6 | |||||
| Software | LS 128.8 |
For the way straightness and flatness
measurements with the LS 128F, the LS 128A must be used (except
software LS 128.7). The standard configuration for position
measurement except the LS 20.1, LS 21.1, LS 70, LS 90.1, LS 90.3
is used.
| Technical specification |
| Angular Measurement | Flatness Measurement | ||
| Longitudinal range | 20 m | 20 m | |
| Range of measurement | ±10o | ±1.5 mm | |
| Resolution | 0.01 " | 5 nm (for Base 100 mm) | |
| Accuracy | (± 0,1 ±0,1 M)'' | ( ±0,02 M2) µm | |
| M - distance
between the LS 128.1 and the LS 128.2 in m |
|||